Error Messages
5
707
707: Calibration error; cal edge time: rise time cal error
707: Calibration error; cal edge time: fall time cal error
707: Calibration error; cal edge time: default values loaded
Indicates a failure in the rise-time or fall-time circuitry has prevented calibration.
The edge time was calibrated using default values, limiting accuracy. See the
Keysight 33210A Service Guide for service information.
850
Calibration error; setup is invalid
You have specified an invalid calibration setup number with the CAL:SET
command. Refer to the Keysight 33210A Service Guide for more information on
the calibration procedures.
851
Calibration error; setup is out of order
Certain calibration setups must be performed in a specific sequence in order to be
valid. Refer to the Keysight 33210A Service Guide for more information on the
calibration procedures.
Keysight 33210A User's Guide
327