Télécharger Imprimer la page

Keysight 33210A Guide D'utilisation page 114

Générateur de signaux arbitraires/fonctions 10 mhz

Publicité

3
Features and Functions
Burst Mode
You can configure the function generator to output a waveform with a specified
number of cycles, called a burst. The function generator can produce a burst
using sine, square, or ramp waveforms (pulse, noise, dc, and arbitrary waveforms
are not allowed).
For more information on the fundamentals of the burst mode, refer to
"Tutorial".
To Select Burst
– The function generator will not allow burst to be enabled at the same time that
– Front-Panel Operation: You must enable burst before setting up any of the
– Remote Interface Operation: To avoid multiple waveform changes, enable the
114
sweep or any modulation mode is enabled. When you enable burst, the sweep
or modulation mode is turned off.
other burst parameters. Press
settings for frequency, output amplitude, and offset voltage.
burst mode after you have set up the other parameters.
BURSt:STATe {OFF|ON}
to output a burst using the present
Keysight 33210A User's Guide
Chapter 7,

Publicité

loading