4. Operation
The standby button can be used to switch the device on and off. This function corresponds to the
soft start switch of conventional PCs. This does not disconnect the power supply from the mains.
Your device has a capacitive keypad. It is enough to touch the glass pane on the key icon to trigger
the function of the key. To avoid unintentional triggering of the keys during wipe disinfection, the
keys must be pressed for at least 0.5 sec.
To switch off the device, the standby button must be pressed for at least 2.5 sec.
The device confirms all entries with a signal tone.
4.1 Keypad
The device is operated via the front keypad. The assignment of the function keys and LEDs varies
depending on the selected equipment variant.
OEM Dräger Standby Taster, LED seitlich neben Taste
The following key functions are integrated:
Stand 15.12.2020
Function keys
Switch on/off
OEM Dräger Standby Taster, LED seitlich neben Taste
OEM Dräger Standby Taster, LED innerhalb des Schlosses
Shutdown in standby
Stand 15.12.2020
Stand 15.12.2020
Device off;
OEM Dräger Standby Taster, LED innerhalb des Schlosses
disconnected from the mains
Stand 15.12.2020
ohne Normbezug
Lampe aus
Device in standby;
connected to power grid
ohne Normbezug
Lampe blinkt
Device switched on
ohne Normbezug
Lampe an
Illustration: Keypad
10
OEM Dräger Standby Taster, LED seitlich neben Taste
Stand 15.12.2020
LED indicators
Deactivate resp. reactivate the touch
OEM Dräger Standby Taster, LED innerhalb des Schlosses
Stand 15.12.2020
Touch shutdown inactive
Lampe aus
ohne Normbezug
Touch shutdown active
ohne Normbezug
Lampe blinkt
5. Maintenance
5.1 Care and hygiene instructions
The device cannot be sterilized.
Before starting disinfection on the running device, use the touch shutdown key to avoid uninten-
tional incorrect entries.
All surface disinfectants on the RKI list are approved for cleaning (see: "List of disinfectants and
disinfection methods tested and approved by the Robert Koch Institute," Federal Health Gazette
12/2013.) The complete list can also be viewed on the RKI website or requested from ACL.
The following table lists active ingredients that are considered harmless to the device within the
specified limits. The table does not represent an evaluation of the disinfection success from a
medical point of view.
Active ingredient
max. concentration in %
Alcohol
100
Biguanide
8
Chlorine, organic or
10
inorganic substances with
active chlorine
Alkaline solution
8
Per-compounds
8
Phenol or phenol
6
derivatives
The manufacturer's instructions for the disinfectants must be observed. The use of disinfectants
based on other active ingredients takes place at the user's own responsibility.
Please ensure that the device has been disinfected with a suitable cleaning agent before any repair
work is carried out, especially when returning the device for repair. If this is not possible, clearly
mark the contaminated product and seal it twice in safety film.
To protect the health of our employees, we reserve the right to refuse acceptance of contaminated
equipment or to apply appropriate follow-up measures at the expense of the sender.
ALWAYS A STEP AHEAD
max. exposure time in
Example products
minutes
360
Bacillol® 30 Foam
Meliseptol
360
Incidin Plus
360
Incidin perfekt
Optisept
360
Milk of lime
360
Incidin active
Terralin PAA
360
Helipur
EN