18.5 Maintenance & Cleaning
Maintain and clean the ionizer at regular intervals
Clean the ion source:
Exchange ion source:
Cleaning
To clean the housing, do not use aggressive cleaning agents (solvents etc.) instead
clean with a soft cloth soaked in mild soap suds. Take care that the device is not
penetrated by fluids and polish it with a dry soft cloth.
Remove loose sample residues / powder carefully with the help of a brush or hand-
held vacuum cleaner.
To clean the ion source use the delivered cleaning brush or a cotton stick moistened
with alcohol. Take care that the peaks are not bent.
Remove loosely attached dust with compressed air.
TABP-BA-e-2212
After 1000 hours
After 30000 hours
Disconnect the device from mains before cleaning.
Do not dismantle the ionizer.
Clean the ion source with care.
Take care that the peaks are not bent.