8.3.3 Adjustment test with internal weight
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Ensure that the CAL-key is allocated with the <Internal adjustment test> function,
see chap. 8.3.1.
To start the test, press the CAL-key.
The difference to the previous adjustment will
be displayed.
8.3.4 Adjustment with external weight
Ensure that the CAL-key is allocated with the <External adjustment> function, see
chap. 8.3.1.
Press the CAL-key and the weight value for
the adjustment weight will be flashing on the
display. (To change the weight value, press
the MENU button and follow the instructions
on the display*).
Put the required adjustment weight carefully in
the center of the weighing plate. Close wind
screen doors completely. Wait until the
request for removing the adjustment weight is
displayed.
Take away adjustment weight.
*The adjustment weight to be used depends on the capacity of the scale.
Carry out adjustment as near as possible to the balance's maximum weight
(recommended adjustment weight see chap. 1). Weights of different nominal
values may be used for adjustment but are not optimal for technical
measuring. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com
TABP-BA-e-2212
38