8.5 Regular inspections
The ABP series supports regular inspection of your scale. This function may be used
to check repeatability, off-center stress (off-center load errors) and linearity. The
instructions on the display support the implementation of individual steps.
Parameter setting:
Call up menu:
Press and hold the CAL-key for approx.
3 sec until the <CAL key setting> menu
appears.
Use the navigation keys to select
<Periodic inspection> and confirm using
the OK-key.
1. Settings Repeatability
<1. Select <Repeatability inspection>
and confirm using the OK-key.
Select desired setting and confirm using
the OK-key.
To enter the weight value for the test
weight, select <Weight value> and
confirm using the OK-key.
Use the navigation keys to enter the
value and confirm using the OK-key.
Repeat these operating steps to set all
other settings for
<Tolerance> / <Number of repetitions>.
Return to menu by
2. Settings off-center load
Repeat these operating steps for
Settings as shown for "Item 1
Repeatability".
TABP-BA-e-2212
.
40