14.1 Parts counting
With parts counting you can either count parts into a container or remove parts from
a container. To count a greater number of parts the average weight per part has to
be determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness. High reference must be selected for
small parts or parts with considerably different sizes.
14.1.1 Settings
+
Enable function and calculate single weight by weighing a known
reference quantity.
1. Selecting an application
 see chap. 11.1.4.
The available applications will be displayed.
Use the navigation keys to select   <Parts
count>. The framing will indicate the current
selection. Confirm using the OK-key and the
application specific settings will be displayed.
2. Storage location no. / name for initial
input
Use the navigation keys ,  to select a
memory space and confirm using the OK-key.
For the first input the display for entering a
memory name will appear.
Use the navigation keys ,  to select a
memory space and confirm using the OK-key.
If required, change name and confirm using
the OK-key.
To overwrite a stored piece weight continue by referring to chap. 14.1.4
TABP-BA-e-2212