12 Description of individual functions
12.1 Zero setting and tare function
Selectable functions:
1. Zero tracking
+
see chap. 12.2.
2. Auto tare function
see chap. 12.3
12.2 Zero tracking
The icon
1. Call function
see chap. 11.1.2.
or
Press and hold the TARE-key for a long time
Use the navigation keys to select , <Zero
tracking> and confirm using the OK-key.
2. Enable/disable function
Use the navigation keys , to enable (on) /
disable (off) the function and confirm using the
OK-key.
3. Back to weighing mode, press ON/OFF
button.
TABP-BA-e-2212
Description
This function is used to correct automatically small
weight variations which appear directly after
switching-on.
In the event that small quantities are removed or added
to the material to be weighed, incorrect weighing results
can be displayed due to the "stability compensation".
(e.g. slow flow of liquids from a container placed on the
balance, evaporating processes).
When apportioning involves small variations of weight, it
is advisable to switch off this function.
After data output an automatic taring is carried out
will be displayed while the zero tracking function is enabled.
64