8.3.5 Adjustment test with external weight
Ensure that the CAL-key is assigned to the <External Weight Test> function, see
chap. 8.3.1.
Press the CAL-key and the weight value for
the adjustment weight will be flashing on the
display. (To change the weight value, press
the MENU button and follow the instructions
on the display).
Put the required adjustment weight carefully
in the center of the weighing plate. Close
wind screen doors completely. Wait until the
request for removing the adjustment weight is
displayed.
The difference to the previous adjustment will
be displayed.
8.4 Adjustment log
If an optional printer is connected and the GLP function enabled, this will be followed
by automatic log output after every adjustment.
Printout example (KERN YKB-01N):
+ For enabling / defining GLP-function see chap. 15.8.3
39
CAL –INTERNAL
KERN & Sohn GmbH
TYPE ABP 300-4M
SN D319300002
BALID 1234
DATE 2018 Aug. 17
TIME 09.14.21
REF 200.00000 g
BFR 200.00000 g
AFT 200.00000 g
-COMPLETE
-SIGNATURE-
Mode of adjustment
Company
Model
Serial no.
Balance identification no.
Date
Time
Used adjustment weight
Before adjustment
After adjustment
prepared by