8.2 Time-controlled automatic adjustment
With the help of its internal adjustment weight and integrated clock the balance can
be set to carry out an automatic adjustment at set times (up to three times daily). This
function is a very convenient function, when adjustment reports are desired to be
made for regular adjustments, or when wishing span adjustments during break times
to avoid interruption of measurement work.
The weight symbol
before it begins. Automatic adjustment can be stopped by actuating the [ON/OFF]
key during this message.
Parameter setting:
Press and hold the CAL-key for approx.
3 sec until the <CAL key setting> menu
appears.
Use the navigation keys to select <CAL
Timer> and confirm using the OK-key.
Select first time <Timer 1> and confirm
using the OK-key.
Select settings [on] or [off] and confirm
using the OK-key.
To set the time select [Settings] and
confirm using the OK-key.
Use the navigation keys to enter time
and confirm using the OK-key.
Repeat these operating steps to set the
time for <Timer 2> / <Timer 3>.
TABP-BA-e-2212
blinks for about two minutes as notification of span calibration
36