Télécharger Imprimer la page

KERN&SOHN ABP Serie Mode D'emploi page 183

Publicité

Les langues disponibles

Les langues disponibles

7.4.1 Turning On the Power
Models d = 0.0001 g
Supply power to balance via mains
adapter. The display lights up and the
balance carries out a selftest. Internal
adjustment will be started automatically
(See chap.8.3.2). The motor noise of the
loading system for the internal
adjustment weight will be audible.
You can cancel the adjustment by
pressing the ON/OFF key.
The selftest is completed when „OFF"
appears on the display. From that point
onwards the weighing scale will be in
standby mode. The weighing balance will
remain switched on as long as it is
connected to the power supply.
To turn the display on/off, press the
ON/OFF key.
When the log-in function is enabled, use
the navigation keys to select the
respective user and enter password, see
chap. 12.8
7.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap.1). During this
warming up time the balance must be connected to the power supply (mains,
rechargeable accumulator or battery). The accuracy of the balance depends on the
local acceleration of gravity. Strictly observe hints in chapter Adjustment.
TABP-BA-e-2212
Models d = 0.00001 g
32

Publicité