Prevent damage caused by dropping, vibration or shock; see sticker on
the left side.
Always use the genuine power pack. The stated voltage value must be
the same as the local voltage.
Risk of injury! The peaks of the ion source are very sharp and cutting.
The ionizer generates poisonous ozone; ensure sufficient ventilation.
For maintenance and repair work disconnect the ionizer from the mains.
Disconnect the ionizer from the mains during periods of idleness.
CAUTION
Maintain and clean the ionizer at regular intervals
Clean the ion source:
After 1000 hours
Exchange ion source:
After 30000 hours
Starting up a damaged ionizer may result in a short circuit, fire or electric
shock.
Starting up out of doors and inside vehicles is prohibited and will result in
total loss of warranty.
The occurrence of electromagnetic fields may result in major display
deviations (incorrect weighing results). Discharge sample at a sufficient
distance from the weighing balance.
During normal operation the green LED [POWER], will be glowing, in
case of a breakdown the red LED [ALARM].
If the red LED is glowing, turn off the ionizer at the master switch and
turn it on again. If the red LED continues to glow, inform the
manufacturer.
During ionization the blue LED [RUN] is lighting.
During ionization an operation noise is audible.