Settings
1. Call up weighing settings
In weighing mode press MENU button.
Press the R-Taste key and use the navigation
keys , to select [
and confirm using the OK-key.
2. Activate function
Use the navigation keys , to select
<Pass/fail evaluation> and confirm using the
OK-key.
Select Settings <On> and confirm using the
OK-key.
3. Setting limits
Select <Setting> and confirm using the
OK-key.
Define the limit values one by one and
confirm using the OK-key.
When entering the limit values ensure that
the values match logically one with
another, i.e. the lower limit value must not
be greater than the upper one.
To return to control mode, press the
ON/OFF-key
TABP-BA-e-2212
Weighing Settings]
118