8.3 Manual adjustment via key [CAL-key]
8.3.1
Setting adjustment function for CAL-key
It is possible to start the preset adjustment procedure without having to access the
menu. The Set Adjustment procedure may be set by simply pressing the [CAL]-key
when in weighing mode.
Press and hold the CAL-key for approx. 3
sec until the <CAL key setting> menu
appears.
Confirm using the OK-key and the available
settings will be displayed.
Adjustment with internal weight, see chap. 8.3.2
For adjustment test using internal weight see
chap. 8.3.3
Adjustment with external weight, see chap. 8.3.4
Adjustment test with external weight, see chap.
8.3.5
Select Settings with the help of the navigation
keys and confirm using the OK-key.
Use the ON/OFF button to return into
weighing mode
8.3.2 Adjustment with internal weight
Ensure that the CAL-key is allocated with the <Internal adjustment> function, see
chap. 8.3.1.
Press CAL key, adjustment is started.
After successful adjustment the balance
automatically returns to weighing mode.
In case of an adjustment error (e.g. objects
on the weighing plate) the display will show
an error message, repeat adjustment.
37