04.1 Safety requirements during installation
�
WARNING: Do not expose the apparatus to direct sunlight or to sources of UV light.
�
WARNING: The apparatus is transportable, however it must be handled with care when it is moved.
To ensure that the device is in working order, it is necessary to check the following:
1
Supply of water to the turbine spray;
2
Supply of air to the turbine;
3
Presence of powder;
�
WARNING: If the supply of water and air is sufficient for the turbine, the Easyjet must work properly.
Standard supplies from the dental unit for a turbine:
-
water from 0 to 3 bars (from 0 hPa to 3000 hPa), max. 40°C;
-
air 2.3 – 3.2 bars (2300 hPa - 3200 hPa)
04.2 Connection of the device
�
WARNING: Check that the connector of the turbine is in good condition.
�
WARNING: Do not connect the Easyjet pro when the turbine is under pressure.
�
WARNING: Do not press the turbine control pedal while connecting the Easyjet pro.
�
WARNING: If your turbine has a light, turn the light off.
�
WARNING: The front polisher terminal and the fitting of the turbine must be perfectly dry. The
presence of moisture could cause clogging of the channels in which the air/powder circulate.
The Easyjet pro has a specific adaptor corresponding to the fitting of the turbine on your dental unit. Use
the Easyjet pro only for the fitting on your own turbine. Fitting the Easyjet pro to a different connector
could damage it.
26
easyjet pro - manual of use and maintenance
04.0 Installation