Product Line for the
Selectra CS Lead Introducer System
The Selectra CS lead introducer system consists of the accessory kit described in this
technical manual and a selection of compatible guiding catheters with different curve
shapes and working lengths. Table 2 shows our product line.
Figure
Item
Outer guiding catheters for Selectra
Selectra Amplatz 6.0-45
Selectra Amplatz 6.0-55
Selectra BIO 2-45
Selectra BIO 2-55
Selectra Extended Hook-45
Selectra Extended Hook-55
Selectra Hook-45
Selectra Hook-55
Selectra MPEP-45
Selectra MPEP-55
Selectra Multipurpose
Hook-45
Selectra Multipurpose
Hook-55
Selectra Right-45
Selectra Right-55
Selectra Straight-45
Selectra Straight-55
Inner guiding catheters for Selectra
Selectra IC-50-59
Selectra IC-50-69
Selectra IC-90-59
Selectra IC-90-69
Selectra Accessory Kit
Slitter Tool
Suitable accessories
Suitable lead introducer set including valve and side port with an
internal diameter of at least 9 F
Steerable EP catheter ViaCath NG 4/S/5mm (not available in the
U.S) (Only for outer catheters)
Venogram balloon catheter Corodyn P1(not available in the U.S)
(Only for outer catheters)
Table 2: Available components and accessories for the Selectra CS lead introducer
system
Compatible leads and accessories
The individually available Selectra guiding catheters with a nominal inner diameter of
7 French (outer catheter) and 5 French (inner catheter) can only be used in combination
with the accessory kit of the Selectra CS lead introducer system.
Outer guiding catheters and implantation accessories of the Selectra CS lead intro-
ducer system are suited for leads with polyurethane coatings or coatings with compa-
rable anti-friction properties and a diameter between 4.6 and 5.8 F. 
Leads with a diameter of less than 5.0 F can be inserted directly through the inner
guiding catheter.
Suitable CS guiding catheters and combinations of inner and outer catheters for dif-
ferent lead lengths can be found in the following table.
BIOTRONIK CS leads
Outer diameter
Length
l
77 cm
5.8 F
Order number
87 cm
<5.0 F
77 cm
375 519
375 520
87 cm
375 523
97 cm
375 524
Table 3: Compatibility of CS guiding catheters and leads
375 527
Technical Data for the Guiding Catheters
375 528
All guiding catheters
375 529
Material
375 530
X-ray contrast
Radiopaque marker at 
the distal end
375 531
Soft tip length:
375 532
Outer guiding catheters
375 533
Working length
Total length
375 534
Inner diameter
Outer diameter
375 535
Handle
375 536
375 537
Inner guiding catheters
375 521
Working length
Total length
392 289
Inner diameter
392 290
Outer diameter
392 291
Handle
392 292
375 518
383 119
Dilator (only for outer catheters)
Working length
Material
Outer diameter
351 197
Inner diameter
Proximal end
336 074
Table 4: Technical data of the guiding catheters
BIOTRONIK's Selectra CS lead introducer system (including all individual accessories
available separately), used in conjunction with devices for implantation in the coronary
sinus, has been qualified, manufactured, and tested according to proven and accepted
standards and procedures. Nonetheless, the physician must be aware that the acces-
sories may easily be damaged through improper handling or use. Except as set forth in
BIOTRONIK's limited warranty, BIOTRONIK makes no expressed or implied warranties
for its accessories.
Selectra guiding catheter
Outer catheter
Inner catheter
Working length
Working length
45 cm
-
55 cm
-
45 cm
-
55 cm
-
45 cm
-
55 cm
-
45 cm
59 cm
55 cm
-
45 cm
59 cm
69 cm
55 cm
69 cm
Shorter models
Longer models
PEBAX
Barium sulfate in the entire catheter area
Tungsten
(black marker zone, 20 mm wide)
1.0 mm (blue)
45.0 cm
55.0 cm
48.3 cm
58.3 cm
2.44 mm (7.3 F)
2.91 mm (8.7 F)
Ergonomic design, side port with Luer lock connector,
main port with integrated hemostatic silicone valve
59.0 cm
69.0 cm
62.9 cm
72.9 cm
1.85 mm (5.55 F)
2.31 mm (6.93 F)
Ergonomic design, side port with Luer lock connector,
main port with integrated hemostatic silicone valve
and with Luer lock connector
53.5 cm
63.0 cm
PE
2.3 mm (6.9 F)
1.02 mm (3.06 F)
Luer lock connection
Disclaimer
4
Legend for the Label
Manufacturing date
Use by
Temperature limit
BIOTRONIK order number
Lot number
Sterilized with ethylene oxide
Do not resterilize
2
STERILIZE
Do not reuse
Non-sterile
NON
STERILE
Follow the instructions for use
Contents
Do not use if packaging is damaged
CE mark
Inner diameter
Outer diameter
Total length
Working length
Guide wire
Torque tool for OTW guide wires
Dilator
Selectra guiding catheters (various curve
shapes)
Selectra transvalvular insertion tool (peel-
away introducer TVI)
Slitter tool (Selectra)
Selectra one-way stopcock
Selectra one-way valve
Selectra sealing cap
Syringe
Caution:
Federal (U.S.A.) law restricts this product to
sale by, or on the order of, a physician.

Publicité

Chapitres

loading