WARNINGS
7.1
The equipment must not be used with any of its parts disassembled: before using the
machine make sure that all its components are properly in place.
For your safety, make sure you pay great attention to the instructions given below.

SAFETY DEVICES

7.2
This equipment includes the following safety devices:
1.
Three micro-switches controlling the door and the automatic locking system: they are
independent of each other and ensure that the door is closed and properly locked. If problems
arise, the user is warned by an alarm and the next cycle cannot start. If any problem arises while
the cycle is running, the microprocessor stops the process and immediately reduces the machine
pressure level.
2.
Two different mechanical thermostats ensure that the temperature of the various components
does not accidentally exceed the preset value. Thermostats are manually reset.
3.
Four electronic temperature sensors continually monitor all critical points on the machine, to
prevent over temperature errors during the operating process.
4.
A pressure relief safety valve protects against the risk of explosion.
5.
An electronic pressure transducer checks all the solenoid valves and opens them if an
overpressure is detected.

FIRST TIME START-UP (USING THE SYSTEM FOR THE FIRST TIME)

7.3
After installing the autoclave (as per paragraphs 6.5 and 6.5.1), ), make sure that all water
connections are watertight. Proceed as detailed below:
1. Start the equipment using the main switch.
2. Enter the USER MENU (paragraph 6.9) and configure the system in accordance with the selected
filling mode (if water filling is from the water mains, check that inlet water connections
are as per paragraph 6.5.1).
The machine is delivered without any water inside
3. Fill the tanks with water: Once the clean water LIV MAX (MAX. LEVEL) icon is displayed, the
test cycle can start (paragraph 7.3.2).
4. Press DOOR to open: insert at least the tray holder and start the cycle by using the START button:
the autoclave will run automatically until the end of the cycle, when a warning sound will be heard
and the display will show "END CYCLE".
160 of 228
CHAPTER 7: USING THE EQUIPMENT
MAINTENANCE AND USE MANUAL
EN
zetaclave B