ERRORS
13.2
The following table contains a list of messages which may appear on display:
m
essage
At the time of power-up you are
OPEN THE DOOR
prompted to open the door so that
the machine can perform a pressu-
re check-up.
A cycle has been started with the
door open.
DOOR OPEN
Even though the door is perfectly
DOOR NOT
closed, it is read as "not locked" due
LOCKED
to mechanical reasons (if this error
persists, contact technical support).
A cycle has been started with the
TO LOAD WATER
clean water minimum level message
flashing on the display.
A cycle has been started with the
TO DRAIN WATER
dirty water maximum level message
flashing on the display.
Attempt to fill the tank with clean
FULL WATER
water with the full tank icon already
RESERVOIR
present on the display.
Door lock system fails to open
completely when cycle terminates.
DOOR STILL
LOCKED
Attempt to start a VACUUM TEST
CHAMBER
cycle with a chamber temperature
TEMP.>40°
higher
temperatures
starting the test cycle.
The autoclave cycle counter prompts
the operator to check water
REPLACE
purifying system resin quality.
RESINS
The machine has run an excessively
SERVICE
large number of cycles and needs
periodic overhauling by a technician.
172 of 228
C
ause
than
40°C;
wait
to
drop
before
MAINTENANCE AND USE MANUAL
EN
s
olution
Open the door to allow automatic
setting of atmospheric pressure (two
beeps).
Close the door and start the cycle.
Press the DOOR button, open and close
the door, then press the START button
to start the cycle.
Fill the tank with clean water
(paragraph 8.2).
Drain dirty water from the tank
(paragraph 9.2).
Disconnect the water hose. It is now
possible to start a new cycle.
Close the door again and start a new
cycle: stop the cycle after a few secon-
ds
with
the
button (paragraph 3.3.1), ), reset the
alarm (paragraph 13), then try
opening the door with the DOOR but-
ton. If the error persists, contact techni-
cal support.
Wait for the temperature on display to
go down below 40° with the door
open; then the test cycle can be started
for
If everything is OK, keep the "LOAD
WATER" button pressed for 5'' (until the
sound sets off ).
Contact your dealer as soon as possible
and request the services of a technician.