ALARMS
13.1
C
oDe anD meaning
AL0001
Cycle stopped by user.
AL0002
No mains supply
AL0003
Door open during cycle
AL0004
Timer stopped
AL0011
No 1st vacuum
AL0012
No 2nd vacuum
AL0013
No 3rd vacuum
AL0015
No vacuum when drying
AL0021
No 1st rise
AL0022
No 2nd rise
AL0024
No final rise
AL0031
No 1st pressure release
AL0032
No 2nd pressure release
AL0034
No final pressure release
AL0100
T1 probe encoding error
AL0101
OPEN T1
AL0102
T1 short circuit
AL0110
High temp. in T1 probe
AL0111
Low temp. in T1 probe
during sterilisation
AL0200
T2 probe encoding error
AL0201
OPEN T2
AL0202
T2 short circuit
170 of 228
p
roblem
Occurs if the START/STOP button is pres-
sed for 1 second or longer
Caused by voltage peaks higher than
10% or by power failures.
Occurs if a control micro-switch detects
an open door while the cycle is running.
The timer will stop when PCB battery
power is insufficient.
The alarm is triggered only if the first
vacuum is not reached.
The alarm is triggered only if the second
vacuum is not reached.
The alarm is triggered only if the third
vacuum is not reached.
Occurs if the preset vacuum degree is
not reached during drying. If this alarm
is triggered, the machine has already
completed the sterilisation.
The machine is unable to reach the 1st
pressure value set.
The machine is unable to reach the 2nd
pressure value set.
The machine is unable to reach the ope-
rating pressure.
After reaching the first set pressure, the
machine triggers an alarm.
After reaching the second set pressure, the
machine triggers an alarm.
The machine fails to release pressure
while drying.
Alarm due to PCB self-diagnosis.
System detects that T1 probe is open.
System detects short-circuit in T1 probe.
T1 probe has exceeded the preset cycle
temperature.
Temperature detected by T1 probe
during sterilisation is below the
minimum set value.
Alarm due to PCB self-diagnosis.
System detects that T2 probe is open.
System detects short-circuit in T2 probe.
MAINTENANCE AND USE MANUAL
EN
s
olution
Reset alarm =
START/STOP + SELECT for 5
Reset the alarm and repeat the cycle.
Reset the alarm and repeat the cycle. If the pro-
blem persists, contact technical support.
Reset the time and date again following the
instructions in paragraph 6.9.2 and 6.9.3). Leave
the autoclave ON for at least one hour.
Reset the alarm, clean the door gasket as
described in point 14.1, dry the instruments to be
sterilised, make sure the pipes are not obstructed
by bags in the chamber and then repeat the
cycle.
Reset the alarm and repeat the cycle. Check if the
problem persists in the following cycles: in this
case contact technical support.
Reset the alarm, fill the thank until the
maximum icons appears on screen and repeat
the cycle: if the problem persists in the
following cycles, contact technical support.
Reset the alarm, remove the tray holder, clean the
sterilisation chamber inside and repeat the cycle.
Reset the alarm, switch the machine off and on: if
the problem persists, contact technical support.
Reset the alarm then wait for 10 minutes with the
door
open.
Repeat
the
problem continues contact the after sales
service.
Reset the alarm, switch the machine off and on: if
the problem persists, contact technical support.
.
seC
cycle:
if
the
zetaclave B