Automatic Draining Of The Dirty Water Tank; Draining And Cleaning The Clean Water Tank; Water Quality Table (Din En 285) - Zhermack Zetaclave B 18 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

MAINTENANCE AND USE MANUAL
To eliminate the need to empty the dirty water tank manually, use the tap at the rear of the autoclave.
Connect a hose to the tap, secure it with a metal hose clamp, and insert then the other end of the hose
into a container or connect it directly to a water drain (paragraph 6.5.1). Leave the tap open: the dirty
water will be drained automatically into the collection container or into the direct drain.
In order to empty the tank and gain access for cleaning purposes, proceed as indicated in
paragraph 6.9.8. This operation should be performed only after disconnecting the water supply line from
the mains (if the sterilizer is connected to an external water purifier).
CEN STANDARD DIN EN 285
Evaporation residue
Silicon oxide (SiO
)
2
Iron
Cadmium
Lead
Heavy metal residues (except iron, cadmium and lead)
Chloride
Phosphate
Conductivity (at 20°C)
pH value
Look
Hardness
The use of water containing contaminants at concentrations exceeding those given in
the table can considerably shorten the operating life of the equipment, seriously
damaging its components, in particular way to the vaporizer, and invalidating the
manufacturer's warranty.
zetaclave B
EN

AUTOMATIC DRAINING OF THE DIRTY WATER TANK

DRAINING AND CLEANING THE CLEAN WATER TANK

WATER QUALITY TABLE (DIN EN 285)

Max value
10 mg/l
1 mg/l
0,2 mg/l
0,005 mg/l
0,05 mg/l
0,1 mg/l
2 mg/l
0,5 mg/l
15 µs/cm
da 5 a 7
colourless, clean, without sediment
0,02 mmol/l
165 of 228
9.2
9.3
9.4

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Zetaclave b 23C306550C306552

Table des Matières