PACLITAXEL ELUTING PERIPHERAL BALLOON DILATATION CATHETER,
- Keep the pressure for 30-60 seconds for an optimal release of the drug. Most of the drug is released
in the first 30 seconds of inflation but, in order to optimize the dilation of the lesion, it is possible to
use longer inflation times greater than 1 minute at the operator's discretion.
- Draw the plunger on the inflation device back to deflate the balloon. Keep up the negative pressure
for between 20 and 60 seconds depending on the size of the balloon. Make sure the balloon is fully
deflated (by means of fluoroscopy) before moving the catheter.
- With negative pressure in the inflation device and with the guide wire in position, carefully retract
the catheter until it can be removed through the introducer.
- Inject contrast liquid through the introducer to check the treated lesion before removing the guide
wire.
- It is recommended that double antiplatelet therapy be administered for a minimum of four weeks
after the intervention, and the treatment may be prolonged at the operator's discretion.
9. Warranty
The device and all its components have been designed, manufactured, tested and packed with all precau-
tionary measures. LVD Biotech guarantees the product up to its expiry date providing its packaging is not
broken, tampered with or damaged.
STANDARDIZED SYMBOLS FOR USE IN LABELLING MEDICAL DEVICES
Catalogue reference
Manufacturer
Consult instructions
for use
6
FOR 0.018" GUIDEWIRE
(EN 980:2008 STANDARD)
Batch no.
Sterilized with ethylene
oxide
2
2
STERILIZE
STERILIZE
2
Do not re-use
STERILIZE
Expiry date
Limitation in storage
temperature
2
2
2
2
STERILIZE
STERILIZE
STERILIZE
STERILIZE
STERILIZE
STERILIZE
2
2
Do not re-sterilize
STERILIZE
STERILIZE
STERILIZE
luminor 18
Keep away from sunlight
Keep dry
2
2
2
Do not use if the pack is
damaged