800
700
600
500
400
300
200
100
0
1.E-05
Figure 5
Graph of nitrogen pumping speed vs inlet pressure

Output Power during High Temperature Condition

When the body temperature reaches a value >45 °C, the maximum output
power decreases with a rate of 3 W/°C up to a max body temperature of 55 °C.
If the body temperature exceed 55 °C the controller goes in status 6 (Fail).
Turbo-V 903 Pumping System User Manual / 87-901-016-01
Nitrogen Pumping Speed
1.E-04
1.E-03
Inlet Pressure [mbar]
Technical Information
TV 903 Pumping System Outline
1.E-02
(A)
18
1.E-01
267/306