Technical Information
Output Power during High Temperature Condition (derating)
Output Power during High Temperature Condition
(derating)
An enhanced power derating is implemented to provide the following
benefits:
The enhanced power derating dynamically adapts per the following
functioning parameters:
WARNING!
The user must set the gas type load (WIN 157) and Cooling mode (WIN 106)
before starting the pump.
If the body temperature exceeds 60 °C, the controller goes in status 6
(Fail).
294/334
Immediate auto-detection of changing requirements in the
application.
Dynamic TMP performance adapting to application conditions for
customer's process stabilization and speeding-up.
Always the best "tuning" for TMP pumping technology taking the
complete advantage of TwisTorr technology potential.
Gas type load (selectable WIN 157: see the "Window Meanings"
paragraph in the relevant controller manual).
Cooling mode (selectable WIN 106: see the "Window Meanings"
paragraph in the relevant controller manual).
Actual pump temperature (can be read by WIN 204: see the
"Window Meanings" paragraph in the relevant controller manual).
Actual Driving Frequency (can be read by WIN 203: see the
"Window Meanings" paragraph in the relevant controller manual).
TwisTorr 704 FS User Manual / 87-901-051-01 (A)