D.1.8 Dc Current (Calys 100); D.1.9 Resistance (Calys 100) - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALYS 50 / 75 / 100
- Range 10V : < (6 ppm R + 1 ppm Cal) /°C,
- Range 50V : < (5 ppm R + 2 ppm Cal) /°C.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.1.8
DC Current (CALYS 100)
Range
Resolution
±50 mA
1 µA
4-20 mA
1 µA
0-20 mA
1 µA
Temperature coefficient: < (10 ppm R + 2 ppm Range)/°C from -10°C to +18°C and from +28°C to +50 °C.
Loop power supply = +24 V typ. (30 V max)
HART compatibility: input impedance Rin = 280 Ω.
Linear or square law display scale.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.1.9
Resistance (CALYS 100)
Range
Resolution
400 Ω
1 mΩ
4000 Ω
10 mΩ
Temperature coefficient: < (6 ppm R + 1 ppm Range)/°C from -10°C to +18°C and from +28°C to +50 °C.
Automatic detection of connection scheme: 2 wire, 3-wire or 4-wire.
For 2 wire connection, the measurement includes the resistance of the line.
For 3 wire connection, add the out-of-balance of the line resistances.
Open circuit terminal voltage < 10 V.
Continuity test:
Open circuit for R > 1000 Ω.
Closed circuit for R < 1000 Ω.
154 / 164
NTA47227-300A8
Accuracy / 1 year
0.012% R + 2 µA
0.012% R + 2 µA
0.012% R + 2 µA
Accuracy / 1 year
0.010% R + 10 mΩ
0.010% R + 100 mΩ
TECHNICAL SPECIFICATIONS
Notes
Rin < 25 Ω
Rin < 25 Ω
Rin < 25 Ω
Notes
Measurement current = 0.25 mA
Measurement current = 0.25 mA