CALYS 50 / 75 / 100
Use the following buttons to define the points.
To edit an existing calibration point,
To add a calibration point,
To delete a calibration point.
Use the F1 (
) button to select the transmission mode field, press the function button, F4 (
the menu and use the Up
Definitions of transmission modes:
Manual: the settings are generated under manual control with the user accepting each point,
One-Way: the settings are generated automatically in the order in which the points (Point 1, Point 2, Point3 etc.) are
defined,
Two-Way: the settings are generated automatically in the order in which the points (Point 1, Point 2, Point3 etc.) are
defined and then in reverse order (Point n, Point (n-1) etc.).
The settling time field can be used to define the time, in seconds, between sending the setting from the output of the
CALYS 50 / 75 / 100 and making the measurement at its input.
When all fields have been completed, press ENTER.
142 / 164
NTA47227-300A8
/ Down
arrow buttons to choose the transmission mode.