D. TECHNICAL SPECIFICATIONS
The accuracies quoted apply at + 18°C to + 28°C unless otherwise stated, and are expressed as ± (n % R + C) where R =
The reading and C = a Constant expressed in practical units. The specifications are given for a confidence level of 95%.
They apply to a product placed under reference conditions of measurement defined hereafter:
- Power up unit and warm up for 30 minutes.
- Using instrument not connected to mains (230 V or 110 V).
- Wait for thirty minutes after disconnecting AC/DC power adaptor.
- For weak signals (100 mV / 1 V cal voltage measurement and simulation and Ohm measurement and simulation) using
connections with Y-shaped lugs or bare wires.
- Acquisition in filtered mode (averaged)
The accuracy includes the accuracy of the reference standards, non-linearity, hysteresis, repeatability and long term stability
over the period quoted.
D.1 Measurement Function (Channel 1 and Channel 2)
DC Voltage
Range
Measuring range
+100 mV
- 10 mV to + 100 mV
+1 V
- 100 mV to + 1 V
+10 V
- 1 V to + 10 V
+50 V
- 5 V to + 50 V
(1) WEM41012-00B 10 MΩ; WEM41012-00A 1 MΩ;
Temperature coefficient < (5 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
DC Current
Range
Measuring range
0-20 mA
0 mA to 24 mA
4-20 mA
3 mA to 24 mA
100 mA
0 mA to 100 mA
Temperature coefficient < (6 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.
Loop power supply = 24 V typical (30 V max)
HART compatibility: input impedance Rin = 280 Ω
Linear or square law display scale.
These specifications are provided for the following configurations:-
- CALYS 150 active mode (+24 V ON)
- CALYS 150 passive mode (+24 V OFF)
Use the absolute value of the value measured (|L|) to calculate the accuracy.
Resistance
Range
Measuring range
400 Ω
0 Ω to 400 Ω
3600 Ω
0 Ω to 3600 Ω
50 KΩ
0 Ω to 50 KΩ
(1) Range on channel 1 only.
Temperature coefficient < (5 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.
Automatic detection of connection scheme: 2 wire, 3 wire or 4 wire.
For 2 wire connection, the measurement includes the resistance of the line.
For 3 wire connection, add the out-of-balance of the line resistances.
Open circuit terminal voltage: < 10 V.
Continuity test:
Open circuit for R > 1000 Ω
Measurement Function (Channel 1 and Channel 2)
Resolution
1 µV
10 µV
100 µV
1 mV
Resolution
0.1 µA
0.1 µA
0.1 µA
Measurer passive mode (+24 V OFF).
Measurer active mode (+24 V ON)
Resolution
Accuracy / 1 year
1 mΩ
0.006% R + 8 mΩ
10 mΩ
0.006% R + 50 mΩ
100 mΩ
0.008% R + 1 Ω
Accuracy / 1 year
Notes
0.005% R + 2 µV
Rin > 1 GΩ
0.005% R + 8 µV
Rin > 1 GΩ
0.007% R + 80 µV
Rin = 10 MΩ (1)
0.007% R + 0.5 mV
Rin = 10 MΩ (1)
Accuracy / 1 year
Notes
0.007% R + 0.8 µA
Rin < 30 Ω
Rin < 30 Ω
0.007% R + 0.8 µA
0.009% R + 2 µA
Rin < 30 Ω
Notes
4-wire measurement
4-wire measurement
4-wire measurement