B.3.2 Current Generation - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

B.3.2
Current generation
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Idc measurement function, then the range using the function and navigation keys.
Confirm with ENTER.
Depending on the range selected, several transmission modes are available:
Range
24 mA
Resolution
1 µA
Loop power supply
No
Set to scale
No
Connection is made between the mA and COM terminals.
If the loop power supply is on, the CALYS 50 / 75 / 100 simulates a passive transmitter supplied with 24 V internally.
If the loop power supply is off, the CALYS 50 / 75 / 100 simulates a passive transmitter supplied with 24 V externally.
When the square law scale is selected, it must be activated by using the Configuration
to scale" is activated, the user enters the values to be simulated in a unit of the new scale.
The CALYS 50 / 75 / 100 indicates the configuration selected in the window, using the following icons:
: To show Loop power is off
: To show Loop power is on
: To indicate a square law scale
Refer to paragraph B.2.2 (Current measurement) for an explanation of the square law scale.
Generation / Simulation
4-20 mA
0-20 mA
1 µA
1 µA
Possible
Possible
Linear or square law
Linear or square law
B
Set to scale menus. Once "set
NTA47227-300A8
121 / 164