B.2.2 Measuring Current - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

B.2.2
Measuring current
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the Idc measurement function then the correct range of measurement using the function and navigation
keys.
Confirm with ENTER.
Depending on the range selected, several modes of measurement are available:
Range
50 mA
1 µA
Resolution
< 30 Ω
Input impedance
Loop power supply
No
Set to scale
No
If loop power supply is on, the connection is made between terminals 4-w and mA.
In this case, the CALYS 50 / 75 / 100 supplies a passive transmitter with 24 V and at the same time measures the current
established by the transmitter.
If the loop power supply is off, the connection is made between terminals mA and COM.
When set to the square law scale, the calibrator takes the square root of the input and displays the result as a percentage.
For example, if the calibrator is connected to the output of a differential pressure transmitter, it displays a result
proportional to the flow rate.
4-20 mA
1 µA
< 30 Ω
< 280 Ω if HART ON
Possible
Linear or square law
Measurement
0-20 mA
1 µA
< 30 Ω
< 280 Ω if HART ON
Possible
Linear or square law
NTA47227-300A8
B
109 / 164

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Calys 75Calys 100

Table des Matières