CALYS 50 / 75 / 100
A.6 Service
The unit must always be set up according to the instructions in this notice. Incomplete or poorly executed setting up may
adversely affect the safety of the operator.
The responsible authority must ensure on a regular basis that factors affecting safety do not change with time and carry
out any necessary preventive work.
Before opening the unit for any work, you must ensure that all wires are disconnected from the unit
Any adjustment, maintenance or repair of an open unit must be avoided as far as possible and, if essential, must be
carried out by qualified personnel who are familiar with the risks involved.
A.6.1
Software updates
The software is updated by the UPG32 program available on the Web-site, www.aoip.com. To find out which version of
firmware is installed in your unit, use the Configuration
The quickest way to find out if an update is available is to visit the AOIP S.A.S website and look at the "Software" page.
To update the firmware, proceed as follows:
1) If necessary, install on the PC the USB driver for communication with AOIP instruments. This driver can be
downloaded from our site, along with an information page describing the installation procedure
2) Disconnect the leads connected to the measurement and simulation terminals
3) Connect the instrument to the PC using the USB lead supplied with the product.
4) Download and run the firmware update programme.
5) Select the language then the file containing the firmware and download in the first stage.
6) Choose the communication parameters that match the parameters of the CALYS 50 / 75 / 100. The
communication port used is a virtual port which does not correspond to a physical port on your computer. The
other parameters to be selected are defined in the diagram below.
102 / 164
NTA47227-300A8
About menu.
GENERAL