CALYS 50 / 75 / 100
B.2.9
Measurement by Thermocouple (Temperature)
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the Tc measurement function, then the appropriate "type of thermocouple" using the function and
navigation keys.
Select the display unit
Select the type of cold junction (CSF) used. Enter the temperature of the cold junction in the case of a
programmed cold junction.
Confirm with ENTER.
The thermocouples available are: K, T, J, E, N, U, L, S, R, B, C, PL, Mo, NiMo/NiCo.
After a significant thermal shock, it is recommended that the unit is left for the temperature to stabilize in order to use the
internal cold junction (CSF) with maximum accuracy.