B.2.9 Measurement By Thermocouple (Temperature) - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALYS 50 / 75 / 100
B.2.9
Measurement by Thermocouple (Temperature)
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the Tc measurement function, then the appropriate "type of thermocouple" using the function and
navigation keys.
Select the display unit
Select the type of cold junction (CSF) used. Enter the temperature of the cold junction in the case of a
programmed cold junction.
Confirm with ENTER.
The thermocouples available are: K, T, J, E, N, U, L, S, R, B, C, PL, Mo, NiMo/NiCo.
After a significant thermal shock, it is recommended that the unit is left for the temperature to stabilize in order to use the
internal cold junction (CSF) with maximum accuracy.
116 / 164
NTA47227-300A8
GETTING STARTED

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Calys 75Calys 100

Table des Matières