Advanced Operation; Imulation Modes; C.1.1 Manual Edit Mode - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALYS 50 / 75 / 100
C. ADVANCED OPERATION
C.1 Simulation modes
Several transmission modes are available in the CALYS 50 / 75 / 100 to facilitate rapid checking and calibration of
instruments and transmitters.
To change the transmission mode, open the transmission window using the OUT function key (F2).
When the transmission window is open, the CALYS 50 / 75 / 100 is set by default to the Manual edit mode.
To access the other modes, select the Mode menu using function key F4. Select a transmission mode using the
Up/Down keys of the navigator and confirm with ENTER.
To exit a transmission mode and return to the default mode, press the ESC key.
C.1.1
Manual Edit mode
In this mode, the value to be transmitted may be entered directly using the alphanumeric keys.
The value entered appears at the bottom of the transmission window during entry.
To cancel the entry, press the ESC key. To transmit the value entered, confirm with the ENTER key.
128 / 164
NTA47227-300A8

ADVANCED OPERATION

Setting entered
manually

Publicité