CALYS 50 / 75 / 100
B.3.5
Thermocouple simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Tc transmission function, then the appropriate "type of thermocouple", using the function and
navigation keys.
Select the display unit
Select the type of cold junction compensation (CSF) used. Enter the temperature of the CSF in the case of a
programmed CSF.
Confirm with ENTER.
The thermocouples available are: K, T, J, E, N, U, L, S, R, B, C, PL, Mo, NiMo/NiCo.
After a significant thermal shock, it is recommended that the unit is left for the temperature to stabilize in order to use the
internal cold junction (CSF) with maximum accuracy.
124 / 164
NTA47227-300A8
GETTING STARTED