B.3.5 Thermocouple Simulation (Temperature) - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALYS 50 / 75 / 100
B.3.5
Thermocouple simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Tc transmission function, then the appropriate "type of thermocouple", using the function and
navigation keys.
Select the display unit
Select the type of cold junction compensation (CSF) used. Enter the temperature of the CSF in the case of a
programmed CSF.
Confirm with ENTER.
The thermocouples available are: K, T, J, E, N, U, L, S, R, B, C, PL, Mo, NiMo/NiCo.
After a significant thermal shock, it is recommended that the unit is left for the temperature to stabilize in order to use the
internal cold junction (CSF) with maximum accuracy.