CALYS 50 / 75 / 100
B.3.1
Generating a DC voltage
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Vdc transmission function, then the range suitable for measurement using the function and navigation
keys.
Confirm with ENTER.
The following ranges are available:
Range
100 mV
Resolution
1 µV
Output impedance
< 1 Ω
Output load
1 KΩ
(1) This rating is only available on units whose product code (visible at the back of the unit) starts with WEM41001
The DC source voltage to be generated is connected between terminals V and COM.
120 / 164
NTA47227-300A8
2 V
20 V
10 µV
100 µV
<1 Ω
<1 Ω
2 KΩ
4 KΩ
GETTING STARTED
50 V (1)
1mV
<1 Ω
4 KOhms