B.2.3
Measuring resistance
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the Idc measurement function then the correct measurement range using the function and navigation
keys.
Confirm with ENTER.
The following ranges are available:
Range
400 Ohms
Resolution
1 mOhms
Measurement current
0.25 mA
To carry out a correct resistance measurement with 3 wires, the 3 conductors used must be:
of the same length,
of the same diameter,
of the same type of metal.
The CALYS 50 / 75 / 100 displays an icon showing the connections used (
wire) to make the measurement. The wiring arrangement is automatically detected by the calibrator.
4000 Ohms
Auto
10 mOhms
--
0.25 mA
0.25 mA
Measurement
for 2 wire,
for 3 wire or
NTA47227-300A8
B
4
111 / 164