CALYS 50 / 75 / 100
Continuity test:
Open circuit for R > 1000 Ω.
Closed circuit for R < 1000 Ω.
D.1.4
Temperature by thermocouples (CALYS 50 / 75)
Sensor types:
- In accordance with CEI 584-1/1995 (couples K, T, J, E, S, B, N),
- In accordance with Din 43710 (couples U and L),
- In accordance with the ENGELHARD table (platinum couple).
- As per standard ASTM E 1751-00 (G couple)
- As per standard ASTM E 988-96 (D W3Re/W25Re couple; C W5Re/W26Re couple)
Sensor
Measuring range
- 250 to -200°C
-200 to -120°C
K
-120 to 0°C
+0 to +1,372°C
- 250 to -200°C
-200 to -120°C
T
-120 to -50°C
-50 to +400°C
-210 to -120°C
J
-120 to 0°C
+0 to +1,200°C
- 250 to -200°C
-200 to -100°C
E
-100 to 0°C
+0 to +1,000°C
-50 to +150°C
R
+150 to +550°C
+550 to +1,768°C
-50 to +150°C
S
+150 to +550°C
+550 to +1,768°C
+400 to +900°C
B
+900 to +1,820°C
U
-200 to +660°C
L
-200 to +900°C
-20 to +900°C
C
+900 to +2,310°C
-240 to -190°C
-190 to -110°C
N
-110 to 0°C
+0 to +1,300°C
Platinum
-100 to +1,400°C
Mo
0 to +1,375°C
NiMo/NiCo
-50 to +1,410°C
+ 0 to + 310°C
D
+ 310 to + 1000°C
+ 1000 to + 2000°C
+ 2000 to + 2315°C
+ 0 to + 100°C
+ 100 to + 300°C
G
+ 300 to + 1100°C
+ 1100 to + 1800°C
+ 1800 to + 2315°C
152 / 164
NTA47227-300A8
Resolution
Drift / year
0.2°C
0.80°C
0.1°C
0.25°C
0.05°C
0.1°C
0.05°C
0.013% R +0.08°C
0.2°C
0.70°C
0.05°C
0.25°C
0.05°C
0.10°C
0.05°C
0.013% R +0.08°C
0.05°C
0.25°C
0.05°C
0.09°C
0.05°C
0.013% R +0.07°C
0.1°C
0.45°C
0.05°C
0.15°C
0.05°C
0.07°C
0.05°C
0.013% R +0.05°C
0.5°C
0.8°C
0.2°C
0.013% R +0.35°C
0.1°C
0.013% R +0.2°C
0.5°C
0.80°C
0.2°C
0.013% R +0.35°C
0.1°C
0.013% R +0.25°C
0.2°C
0.013% R +0.4°C
0.1°C
0.013% R + 0.2°C
0.05°C
0.15°C
0.05°C
0.2°C
0.1°C
0.25°C
0.1°C
0.013% R + 00.15°C
0.2°C
0.5°C
0.1°C
0.15°C
0.05°C
0.08°C
0.05°C
0.013% R +0.06°C
0.05°C
0.3°C
0.05°C
0.013% R +0.06°C
0.05°C
0.013% R +0.30°C
0,1°C
0,30°C
0,05°C
0,30°C
0,05°C
0,025% R +0,05°C
0,05°C
0,040% R
0,5°C
2.3°C
0,2°C
0,55°C
0,05°C
0,28°C
0,05°C
0,4°C
0,05°C
0,65°C