B.2.11 Pressure Measurement And Detection Of Contact Opening - AOIP CALYS 50 Mode D'emploi

Calibrateur de process multifonction
Masquer les pouces Voir aussi pour CALYS 50:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALYS 50 / 75 / 100

B.2.11 Pressure measurement and detection of contact opening

The CALYS 50 / 75 / 100 can detect the opening or closing of a contact (switch) and acquire the value of pressure at the
moment that this change of state occurs.
In this case, the pressure measurement function is configured from the transmission part. The transmission/simulation
functions are selected by pressing the function button, F2 (OUT). Pressing this button activates the
transmission/simulation window (see next chapter for more information on the transmission/simulation part).
Open the MEASUREMENT CONFIGURATION dialog box,
Select the Pressure measurement function.
Press ENTER.
To access the Switch Test mode, press the function button, F4, to select the Mode menu and then press ENTER.
Select the type of detection (Open, Close or Open and Close) and press ENTER.
Once the type of detection is selected, every change of the selected type is detected and the measured value of pressure
at the time of change is displayed.
To accept (acknowledge) the measurement, press the function button, F1 (OK).
To exit the Switch Test mode, press ESC.
118 / 164
NTA47227-300A8
GETTING STARTED