The precision is guaranteed for a reference junction temperature of 0 °C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.3 °C at 0 °C. For other
temperatures, account must be taken of the sensitivity of the thermocouple to the temperature (T) in question, giving an
additional uncertainty of 0.3 °C * S(0°C)/S(T).
Temperature coefficient: < 10 % of the accuracy/°C from -10°C to +18°C and from +28°C to +50 °C.
Display in °C, °F and K.
It is possible, thermocouple B excepted, to choose the location of the cold junction by programming from the keyboard:
external at 0°C,
Internal (compensation for the temperature of the terminals of the unit).
By programming the temperature.
D.1.5
Temperature using resistive probes (CALYS 50 / 75)
Sensor
Range of measurement
- 220°C + 1,200°C
Pt 50(α = 3851)
- 220°C + 1,200°C
Pt 100 (α = 3851)
- 200°C + 510°C
JPt 100 (α = 3916)
Pt 100 (α = 3926)
- 210°C + 850°C
- 220°C + 600°C
Pt 200 (α = 3851)
- 220°C + 1,200°C
Pt 500 (α = 3851)
Pt 1 000 (α = 3851)
- 220°C + 1,200°C
- 60°C + 180°C
Ni 100 (α = 618)
Ni 120 (α = 672)
- 40°C + 205°C
- 60°C + 180°C
Ni 1 000 (α = 618)
- 70°C + 150°C
Cu 10 (α = 427)
Cu 50 (α = 428)
- 50°C + 150°C
For negative temperatures, use the displayed value R and not its absolute value.
Temperature coefficient: < 10 % of the accuracy/°C from -10°C to +18°C and from +28°C to +50 °C.
The above accuracy is given for 4 wire connection of the temperature sensor.
Taking into account, also, the intrinsic error of the temperature sensor used and its conditions of use.
Measurement current: 0.25 mA
D.1.6
Frequency and counting (CALYS 50 / 75 / 100)
Range
Resolution
20 kHz
0.01 Hz
Temperature coefficient: < 5 ppm R/°C from -10°C to 18°C and from 28°C to 50°C.
Triggering level 1 V
Scale in beats/min and Hz
Measurement for frequency output and dry contact
In the case of counting, this measurement may be made for a defined time or an infinite time.
D.1.7
DC Voltage (CALYS 100)
Range
Resolution
±100 mV
1 µV
±1 V (1)
10 µV
±10 V
100 µV
±50 V
1 mV
(1) Specification domain: -0.8V to +1V
(2) WEM41012-00B 10 MΩ; WEM41012-00A 1 MΩ;
Temperature coefficient: from -10°C to +18°C and from +28°C to +50 °C.
- Ranges 100mV – 1V : < (6 ppm R + 0.5 ppm Cal) /°C,
Resolution
0.01°C
0.01°C
0.01°C
0.01°C
0,01°C
0.01°C
0.01°C
0.01°C
0.01°C
0.01°C
0.01°C
0.01°C
Accuracy / 1 year
0.005% R + 5 mHz
Accuracy / 1 year
0.010% R + 3 µV
0.010% R + 20 µV
0.012% R + 200 µV
0.012% R + 2 mV
Measurement function
Accuracy / 1 year
0.012% R + 0.06°C
0.012% R + 0.05°C
0.012% R + 0.05°C
0.012% R + 0.05°C
0.012% R + 0.12°C
0.012% R + 0.07°C
0.012% R + 0.05°C
0.012% R + 0.03°C
0.012% R + 0.03°C
0.012% R + 0.03°C
0.012% R + 0.18°C
0.012% R + 0.06°C
Notes
Notes
Rin > 1 GΩ
Rin > 1 GΩ
Rin = 10 MΩ (2)
Rin = 10 MΩ (2)
NTA47227-300A8
D
153 / 164