In measurement mode, channel V1 is connected to the four "IN" terminals located on the left half of the unit:

Measuring DC voltage

Display the MEASUREMENT MENU dialogue box:
Select the Vdc measurement function then the correct measurement range using the function and navigation keys.
Confirm with ENTER.
The following ranges are available:
Range
100 mV
Resolution
1 µV
Input impedance
> 1 GΩ
(1) WEM41012-00B 10 MΩ; WEM41012-00A 1 MΩ;
The DC voltage to be measured is connected between the V and COM terminals.
122 / 194
NTA47224-300A9
1 V
10 V
10 µV
100 µV
> 1 GΩ
10 MΩ
50 V
Auto
1 mV
---
10 MΩ (1)
---
GETTING STARTED