Loading The Digit Power Dispenser; Dispenser Operation - Dentsply Sirona digit Power Dispenser Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

3.7 Loading the digit Power
To test functionality pre-operative, please follow directions below, extruding one cartridge, prior to use on patients.
1. Prepare a digit Power
snapping the circular disk off the cartridge.
4. Ensure the digit Power
into the cartridge cap by the retaining tab.
If cartridge is misaligned (retaining tab not
parallel to cartridge) cartridge cap cannot be
seated onto handpiece.
Improper Assembly – Device Malfunction
• Ensure plungers are fully retracted before placing cartridge cap onto handpiece.
• Ensure the cartridge is fully seated into the cartridge cap.
• Ensure cartridge/cap assembly is fully seated onto the handpiece.
• Audible clicks may be heard to denote proper assembly.
7. Place a protective barrier sleeve over Dispenser.
8. Verify that all parts of the Dispenser are securely attached before use.
9. Charge Status: Handpiece Charge Level must display at least 1 bar. Refer to Section 3.4, Charging the Handpiece.

3.8 Dispenser operation

• The handpiece has two forward speed dispensing buttons for finger operation, operating discrete Slow and Fast speeds.
• The optional wireless foot control allows continuously variable speed control throughout the entire handpiece speed range.
The wireless foot control connects to the handpiece via Bluetooth. Like all wireless devices, the connection is subject to interruption by
environmental interference. If the connection with the handpiece is lost, complete the procedure using the manual speed buttons. When
convenient, re-pair the wireless foot control and handpiece as outlined in Section 3.5.
1. Ensure the integrated timer is properly adjusted for the selected material.
2. Activate the Dispenser by depressing the fast or slow handpiece buttons, or depress the optional wireless foot control and, away from the patient,
bleed a small amount through the tip to assure proper flow and a good mix.
3. Follow Directions for Use for selected material placement.
4. Speed may be changed during operation by repositioning finger from one button to the other. Speed may be varied with the wireless foot control by
stepping harder or softer on the control at any time during operation.
Dispenser
®
cartridge for use by
2. Insert unit dose cartridge into mixing tip.
®
Ensure both cartridge spurs fully lock into the
square holes on the side of the mixing tip.
cartridge is locked
5. Ensure plungers are fully retracted. Visually
®
inspect and confirm plunger retraction before
attempting to seat the cartridge cap.
3. With the cartridge cap removed from the
handpiece, insert the digit Power
mixing tip assembly into the cartridge cap.
6. Place cartridge cap with fully seated
cartridge in place on handpiece so that the
locks slide into place. An audible click will
be heard when the cartridge cap is properly
secured to the handpiece. Tug on the seated
cartridge cap to confirm secure attachment.
cartridge
®
9

Publicité

Table des Matières
loading

Table des Matières