Error Message
MED - 044
The data need to be
synchronised. Press "OK"
as soon as possible.
The process lasts a few
seconds
MED - 047 / MED - 048
Internal problem.
"SHUTDOWN" the device
then start it again.
MED - 049
The data need to be
synchronised.
"SHUTDOWN" the device
then start it again.
MED - 050 a MED -058
Internal problem.
"SHUTDOWN" the device
then start it again.
MED - 063
The irrigation tubing line
shall be filled by pressing
the PUMP button before
using the handpiece.
MED - 065
Handpiece not correct,
change channel or
handpiece.
Table 3 – Error Messages
NOTE: For any error message not listed in
Table 3 at page 46
, contact a Mectron
Service Centre.
46
Actions Required
Complete the current
operation and then run the
synchronization as soon as
possible by pressing the OK
button.
"SHUTDOWN" the device
and then start it again. If the
problem persists, contact
MECTRON assistance.
"SHUTDOWN" the device
and then start it again. If the
problem persists, contact
MECTRON assistance.
"SHUTDOWN" the device
and then start it again. If the
problem persists, contact
MECTRON assistance.
Press the "PUMP" button to
fill the irrigation circuit.
Adopt the handpiece
suitable for the selected
channel.
Description
Problem in the internal data
synchronization.
Internal problem.
Data synchronization
needed.
Internal problem.
Filling cycle performed for
a time shorter than the
minimum required. The
filling cycle that follows
can be interrupted by the
operator.
Handpiece connected, but
not suitable for the channel
or not supported by the
system.