Error Message
MED - 021
Channel out of use.
MED - 022
Internal problem: please
contact assistance.
MED - 028
External power supply
problem. Check the
settings.
MED - 029
Channel out of use.
MED - 030
Internal problem: please
contact assistance.
MED - 032
Lack of internal
communication: please
contact assistance.
MED - 038
"PUMP" button not
active. Fill the irrigation
tubing line by pressing the
footswitch.
MED - 043
Default language - English.
Actions Required
Use the channel that works.
Contact technical assistance.
The device is working
properly: contact the
technical assistance as soon
as possible.
The device is not working
properly: contact
immediately the technical
assistance.
Check that the correctness
of the settings and then
proceed with the operation.
Contact immediately the
technical assistance.
The device is working
properly: contact the
technical assistance as soon
as possible.
The device is not working
properly: contact
immediately the technical
assistance.
The device is working
properly: contact the
technical assistance as soon
as possible.
The device is not working
properly: contact
immediately the technical
assistance.
Fill the irrigation circuit
by means of the foot
pedal, before starting the
intervention.
Select the desired language
by using the "language"
button. If the problem
persists, contact the technical
assistance.
TROUBLESHOOTING
Description
Checksum error, firmware
module corruption.
Checksum error, firmware
module corruption.
Mains power supply
anomaly.
Failure in the execution
of the module interface
program.
Failure in the execution of an
internal process.
Internal communication
failure.
"PUMP" button not available.
Language selection not
properly working: English set
as the default language.
45