Emissions Test
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
36
Compliance
Electromagnetic Environment Guidance
The PIEZOSURGERY® plus is suitable
Class A
for use in all establishments other than
domestic, and may be used in domestic
Class A
establishments and those directly
connected to the public low voltage power
supply network that supplies buildings
used for domestic purposes, provided the
following warning is heeded.
CAUTION: This device is intended for
use by healthcare professionals only.
The EMISSIONS characteristics of the
PIEZOSURGERY® plus make it suitable
for use in industrial areas and hospitals
Compliant
(CISPR 11 class A). If it is used in a
residential environment (for which CISPR
11 class B is normally required) the
PIEZOSURGERY® plus might not offer
adequate protection to radio-frequency
communication services. The user might
need to take mitigation measures ,
such as relocating or reorienting the
PIEZOSURGERY® plus.