Hold the central body of the handpiece
firmly;
CAUTION: Do not grasp the end
and/or cord of the handpiece; only
hold the central body. Do not rotate
the handpiece; simply hold it firmly and
rotate the wrench, only.
Turn the wrench in a clockwise direction
until the clutch engages (the outer part
of the wrench slips, compared to the
handpiece,
producing
clicking sounds "CLICK").
The extension is now properly tightened;
Screw the cover supplied in the kit into
the PIEZOSURGERY
medical+ till the
®
end of the threading;
WARNING: PIEZOSURGERY®
medical+ inserts. Always screw in
the cover supplied with the extension
to protect surrounding tissues from
overheating.
CAUTION: Manually screw the
cover onto the handpiece till the end of
the threading. Do not use any type of
wrench.
20
5
mechanical
a
6
click!