Connect the foot pedal to the back of
the device in the socket marked with
the symbol
, using the foot pedal
cable plug, ensuring it clicks into place.
To disconnect the foot pedal from the
machine body, grasp the connector, crush
the release sleeve and pull backwards;
NOTE: if the foot pedal ceases
functioning, the button on the back, left
of the device, below the symbol
be used to continue treatment.
NOTE: the foot pedal is provided with
a bracket that, if positioned vertically,
enables it to be moved to the most
appropriate place for use, with no need
to touch it with hands.
The bracket can also be positioned
horizontally, if not in use;
CAUTION: The user must pay
particular attention to the foot pedal
positioning, so that the pedal can
only be activated intentionally by the
operator.
Insert the solution bag stand into the
proper holes;
Install the two peristaltic pumps, gently
pushing the head of the peristaltic pump
towards the corresponding base (on the
device). Carefully rotate the pump head
forwards and backwards to ensure the
two parts match;