CAUTION: No changes can be made to
this device.
4.2
Device Installation
Equipotential plug: there is an additional
equipotential plug on the back of the
device. The plug complies with DIN
42801. Insert the equipotential cable
connector (not supplied) onto the plug
on the back of the device. The purpose of
the additional equipotential connection
is to reduce the difference in potential
that may occur during operation between
the device body and the conductive
parts of other objects within the medical
environment;
Insert the power supply cord into its
connector on the back of the device and
then into the wall socket;
CAUTION: Position the device in
such a way as to ensure that the plug
is easily accessible as this is considered
a means for disconnecting the power
supply.
12
WARNING: Before using the
PIEZOSURGERY® medical irrigation kit, check
the integrity of the sterile packaging and
inspect the product to exclude the presence
of any damage. Do not use the irrigation kit
if the packaging is open or damaged. The
irrigation set loses sterility if the packaging
is broken or damaged. If the packaging is
damaged or unintentionally opened before
the use, proceed with the disposal of the set.
The product cannot be re-sterilized or reused.