7.1
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment.
Although PIEZOSURGERY® plus complies
with standard IEC/EN 60601-1-2, it may
interfere with other equipment located
nearby. PIEZOSURGERY® plus must not be
used near, or stacked on other equipment.
Install PIEZOSURGERY® plus at a safe
distance from life-support equipment. If
necessary, however, prior their use, please
check and monitor that the device and all
other equipment works correctly in that
layout.
WARNING: Portable and mobile radio
communication appliances may affect proper
device operation.
7.1.1
Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
PIEZOSURGERY® plus is designed to operate in the electromagnetic environment specified
below. The customer or the user of PIEZOSURGERY® plus should ensure that it is used in such
environment.
Emissions Test
RF Emissions
CISPR 11
WARNING: Contraindications.
Interference by other equipment.
An electrosurgical knife of other
electrosurgical devices near the
PIEZOSURGERY® plus device, may interfere
with its correct operation.
WARNING: The device requires specific
EMC precautions and must be installed
and commissioned according to the EMC
information provided in this chapter.
WARNING: The use of cables and
accessories not supplied by Mectron might
negatively affect the EMC performances.
Compliance
Electromagnetic Environment Guidance
PIEZOSURGERY® plus uses RF energy
only for its internal operation. The
PIEZOSURGERY® plus could emit
frequencies causing interference with other
Group 1
devices in the vicinity.
Install the PIEZOSURGERY® plus away
from life-support equipment. Check correct
operation of all the equipment before
starting the operation.
TECHNICAL DATA
35