Adjustmentandfunctionalities
Flow filter level
slow
(response time 10 to 90% = 14 s)
fast
(response time 10 to 90% = 5 s)
10.6. selecting the measurement
range
The output signals are proportional to the measured fluid velocity.
Switches 4 and 5 are used to adjust the measurement range of the
transmitter to your application.
Select the measurement range by positioning switches 4 and 5
(see Fig. 24 and following table).
Measurement range
0 to 2 m/s
0 to 5 m/s
0 to 10 m/s
0 to full scale calibrated
(between 2 and 10 m/s)
After the measurement range has been modified, the percentages
set for the low and high switching thresholds are applied to the new
full scale selected.
Type 8041
Position switch 3
OFF
ON
Position
Position
switch 4
switch 5
ON
OFF
OFF
ON
OFF
OFF
ON
ON

10.7. calibrating the flow zero point

Calibrate the transmitter on commissioning and after
each maintenance operation.
Before calibrating the zero point on commissioning:
immerse the measuring element in the fluid for 24 h
before calibration.
Before calibrating the zero point after each maintenance
operation:
immerse the measuring element in the fluid for 1 h before
calibration.
Before calibration, ensure that the pipe does not contain
any air bubbles and that the fluid is not moving.
Fill the pipe with fluid.
Stop the flow.
Calibrate the "zero flow" point (see