9.4.1.
Unit – setting the unit for the flow rate
In this parameter, set the unit the measured flow rate is displayed in and in which the limits of the flow rate range
associated to the 4-20 mA output are defined.
When changing the unit (e.g. from liters to gallons), also change the current output limits.
Setting in the parameter:
Status display changes to "active".
The selected unit is shown on the
display.
Figure 11:
Unit – Setting the display for the flow rate unit
9.4.2.
InP (Input) – setting the K factor of the fitting used
→
In this parameter, set the K factor of the fitting used, in pulses/liter.
using a pre-set value (this function is not available on a "Version 1" device. Refer to the nameplate)
When using a pre-set K factor, check that the factor indicated in the instruction manual of the fitting used
is the same as the one displayed by the 8022.
The K factors of the Bürkert sensors types 8020, 8030 and 8070 are pre-set within the 8022, for each form of the
fitting S020, S030 or S070.
→
Choose the sensor type associated to the 8022, then, depending on the type of sensor :
• Type 8020 – choose the fitting, material and DN
• Type 8030 – choose the material and DN
• Type 8070 – choose the DN
The pre-set K factor is displayed.
18
l / s
l / m
l / h
m
3
/ m
m
3
/ h
g / s
g / m
g / s
End
English
Liters/second
Liters/minute
Liters/hour
Cubic meters/minute
Cubic meters/hour
US gallons/second
US gallons/minute
US gallons/hour
Type 8022
SettingandFunctions