Commissioning
L
250 VAC
N
3 A
protective circuit
C
Solenoid valve
NO
(or alarm)
NC
250 VAC, 3 A max.
protective cap on the relay
* If a direct earthing connection is not possible, fit a 100 nF/50V capacitor
between the negative terminal of the power supply and the earth.
Fig. 23 : Relay output connection in normally closed mode
Type 8041
Power supply
18-36 VDC
+ -
*
300 mA
1 2 3 4 5 6
4...20 V+ V- PE Pls+Pls-
8041
9.
commissioninG
9.1.
safety instructions
WarnIng
danger due to nonconforming commissioning.
Nonconforming commissioning could lead to injuries and damage
the device and its surroundings.
•
Before commissioning, make sure that the staff in charge have
read and fully understood the contents of the manual.
•
In particular, observe the safety recommendations and intended
use.
•
The device/installation must only be commissioned by suitably
trained staff.
note
Risk of damage to the device due to the environment
•
Protect this device against electromagnetic interference,
ultraviolet rays and, when installed outdoors, the effects of the
climatic conditions.
When the device is switched on and the cover is open,
protection against electric shock is no longer guaranteed.
English
25