5.

descRipTion

5.1.

Area of application

The 8041 transmitter is used to measure the flow of neutral or slightly
aggressive fluids with a conductivity of more than 20 µS/cm in DN06
to DN400 pipes.
5.2.

General description

5.2.1.

construction

The 8041 transmitter comprises an electronic module and a PVDF
or stainless steel measurement sensor.
The measurement sensor comprises two electrodes and a magnetic
system.
Electrical connection is made via two cable glands by means of a
6-pin terminal block.
The 8041 transmitter requires an 18-36 VDC power supply and has:
one frequency output;
one relay output;
one 4-20 mA current output.
5.2.2.

principle of operation

The magnetic system in the measurement sensor generates a mag-
netic field in the fluid, perpendicular to the flow direction, see Fig. 1 .
The electrodes on the measurement sensor ensure electrical contact
Type 8041
Description
with the fluid. When the fluid flows over them, a voltage is measured
between the two electrodes. This voltage is proportional to the fluid
velocity.
Fig. 1 : Principle of operation of the measurement sensor
5.3.

description of the name plate

1
2
FLOW:8041-FKM -IND LONG SST N=10000
18-36V — /220mA 4-20mA REL:230VAC/3A
552780
11
10
9
8
Fig. 2 :
Name plate on the 8041 transmitter (example)
1.
Process value measured and type of device
2.
Material used for the sensor seal
English
4
3
Compact
W41UE
7
6
5
7